Programmable shunts and headphones: Are they safe together?

Author:

Spader Heather S.1,Ratanaprasatporn Linda1,Morrison John F.1,Grossberg Jonathan A.2,Cosgrove G. Rees1

Affiliation:

1. Department of Neurosurgery, Rhode Island Hospital and Alpert Medical School at Brown University, Providence, Rhode Island; and

2. Department of Neurosurgery, Emory University, Atlanta, Georgia

Abstract

OBJECT Programmable shunts have a valuable role in the treatment of patients with hydrocephalus, but because a magnet is used to change valve settings, interactions with external magnets may reprogram these shunts. Previous studies have demonstrated the ability of magnetic toys and iPads to erroneously reprogram shunts. Headphones are even more ubiquitous, and they contain an electromagnet for sound projection that sits on the head very close to the shunt valve. This study is the first to look at the magnetic field emissions of headphones and their effect on reprogrammable shunt valves to ascertain whether headphones are safe for patients with these shunts to wear. METHODS In this in vitro study of the magnetic properties of headphones and their interactions with 3 different programmable shunts, the authors evaluated Apple earbuds, Beats by Dr. Dre, and Bose QuietComfort Acoustic Noise Cancelling headphones. Each headphone was tested for electromagnetic field emissions using a direct current gaussmeter. The following valves were evaluated: Codman Hakim programmable valve, Medtronic Strata II valve, and Aesculap proGAV. Each valve was tested at distances of 0 to 50 mm (in 5-mm increments) from each headphone. The exposure time at each distance was 1 minute, and 3 trials were performed to confirm results at each valve setting and distance. RESULTS All 3 headphones generated magnetic fields greater than the respective shunt manufacturer's recommended strength of exposure, but these fields did not persist beyond 5 mm. By 2 cm, the fields levels were below 20 G, well below the Medtronic recommendation of 90 G and the Codman recommendation of 80 G. Because the mechanism for the proGAV is different, there is no recommended gauss level. There was no change in gauss-level emissions by the headphones with changes in frequency and amplitude. Both the Strata and Codman-Hakim valves were reprogrammed by direct contact (distance 0 mm) with the Bose headphones. When a rotation component was added, all 3 headphones reprogrammed the Strata and Codman-Hakim valves at 0 mm. At all distances above 0 mm, the headphones did not affect the shunts. The proGAV valve was not affected by headphones at any distance. CONCLUSIONS Although all the headphones studied generated significant gauss fields at distances less than 5 mm, the programmable valve settings only changed at a distance of 0 mm (i.e., with direct contact). Given the subcutaneous location of the valve, the authors conclude that is highly unlikely that commercially available or customary headphones can contribute to the reprogramming of shunts.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

General Medicine

Cited by 9 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3