The PS Medical Strata valve is a programmable shunt valve used in the treatment of hydrocephalus that allows for noninvasive changes in the pressure setting using a magnet. The Strata valve is sensitive to magnetic fields, and reprogramming is frequently necessary after MR imaging. A known but rare complication of the Strata valve is that the rotor can become locked, causing shunt malfunction. This complication can only occur in a first generation Strata valve.