Author:
Strahle Jennifer,Selzer Béla J.,Muraszko Karin M.,Garton Hugh J. L.,Maher Cormac O.
Abstract
Object
The authors investigated the effect of a tablet computer on performance-level settings of a programmable shunt valve.
Methods
Magnetic field strength near the tablet computer with and without a cover was recorded at distances between 0 and 100 mm. Programmable valves were exposed to the tablet device at distances of less than 1 cm, 1–2.5 cm, 2.5–5 cm, 5–10 cm, and greater than 10 cm. For each distance tested, the valves were exposed 100 times to the tablet with the cover, resulting in 500 total valve exposures. The tablet alone, without the cover, was also tested at distances of less than 1 cm for 30 valve exposures. Changes in valve performance-level settings were recorded.
Results
The maximum recorded magnetic flux density of a tablet with a cover was 17.0 mT, and the maximum recorded magnetic flux density of the tablet alone was 7.6 mT. In 100 exposures at distances between 0 and 1 cm, 58% of valves had different settings following exposure. At distances greater than 1 cm but less than 2.5 cm, 5% of valves in 100 exposures had setting changes. Only a single setting change was noted in 100 exposures at distances greater than 2.5 cm but less than 5 cm. No setting changes were noted at distances greater than 5 cm, including 100 exposures between 5 and 10 cm, and 100 exposures of more than 10 cm. For the 30 valve exposures to the tablet without a cover, 20 valve performance-level changes (67%) were noted.
Conclusions
Based on these results, exposure to tablet devices may alter programmable shunt valve settings.
Publisher
Journal of Neurosurgery Publishing Group (JNSPG)
Cited by
18 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献