Abstract
Laser beam drift greatly influences the accuracy of a four degrees of freedom (4-DOF) measurement system during the detection of machine tool errors, especially for long-distance measurement. A novel method was proposed using bellows to serve as a laser beam shield and air pumps to stabilize the refractive index of air. The inner diameter of the bellows and the control mode of the pumps were optimized through theoretical analysis and simulation. An experimental setup was established to verify the feasibility of the method under the temperature interference condition. The results indicated that the position stability of the laser beam spot can be improved by more than 79% under the action of pumping and inflating. The proposed scheme provides a cost-effective method to reduce the laser beam drift, which can be applied to improve the detection accuracy of a 4-DOF measurement system.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献