Affiliation:
1. Institute for Medical Research, Camden, New Jersey 08103
Abstract
A microbiological hood was designed to reduce the danger of airborne infection of laboratory workers. The hood uses absolute filters to deliver sterile air in a laminar flow to the work area. An air curtain across the hood opening permits easy access but separates the worker from aerosols produced in the hood, and protects material inside the hood from contamination by room air. Tests with bacterial and viral aerosols showed that the air curtain is at least 99.96% effective in preventing airborne particles from entering the work area.
Publisher
American Society for Microbiology
Subject
General Pharmacology, Toxicology and Pharmaceutics,General Immunology and Microbiology,General Biochemistry, Genetics and Molecular Biology,General Medicine
Reference11 articles.
1. Air hygiene in dressing rooms for burns or major wounds;Bourdillon R. B.;Lancet,1946
2. Detection and elimination of contaminating organisms;Coriell L. L.;Natl. Cancer Inst. Monograph,1962
3. Dautrebande L. 1962. Microaerosols. Academic Press Inc. New York.
4. Airfiltrationof microbial particles;Decker H. M.;Am. J. Public Health,1963
5. Genetic markers as tracers in cell culture;Gartler S. M.;Natl. Cancer Inst. Monograph,1967