Modified Laminar Flow Biological Safety Cabinet

Author:

Mcgarrity Gerard J.1,Coriell Lewis L.1

Affiliation:

1. Department of Microbiology, Institute for Medical Research, Camden, New Jersey 08103

Abstract

Tests are reported on a modified laminar flow biological safety cabinet in which the return air plenum that conducts air from the work area to the high efficiency particulate air filters is under negative pressure. Freon gas released inside the cabinet could not be detected outside by a freon gas detection method capable of detecting 10 -6 cc/s. When T3 bacteriophage was aerosolized 5 cm outside the front opening in 11 tests, no phage could be detected inside the cabinet with the motor-filter unit in operation. An average of 2.8 × 10 5 plaque-forming units (PFU)/ft 3 (ca. 0.028 m 3 ) were detected with the motor-filter unit not in operation, a penetration of 0.0%. Aerosolization 5 cm inside the cabinet yielded an average of 10 PFU/ft 3 outside the cabinet with the motor-filter unit in operation and an average of 4.1 × 10 5 PFU/ft 3 with the motor-filter unit not in operation, a penetration of 0.002%. These values are the same order of effectiveness as the positive-pressure laminar flow biological safety cabinets previously tested. The advantages of the negative-pressure return plenum design include: (i) assurance that if cracks or leaks develop in the plenum it will not lead to discharge of contaminated air into the laboratory; and (ii) the price is lower due to reduced manufacturing costs.

Publisher

American Society for Microbiology

Subject

General Pharmacology, Toxicology and Pharmaceutics,General Immunology and Microbiology,General Biochemistry, Genetics and Molecular Biology,General Medicine

Reference8 articles.

1. Anonymous. 1973. Effective use of laminar flow biological safety cabinets. National Audiovisual Center (Government Services Administration) Washington D.C.

2. Anonymous. 1974. Laminar flow biological safety cabinet. NIH Specification NIH-03-112a amended Jan. 1974. National Institutes of Health Bethesda Md.

3. Containment of microbial aerosols in a microbiological safety cabinet;Barbeito M. S.;Appl. Microbiol.,1968

4. Biohazard hood to prevent infection during microbiological procedures;Coriell L. L.;Appl. Microbiol.,1968

5. Microbiological studies on the performance of a laminar airflow biological cabinet;McDade J. J.;Appl. Microbiol.,1968

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献