Author:
Pape S.,Fernández García M.,Moll M.,Montero R.,Palomo F.R.,Vila I.,Wiehe M.
Abstract
Abstract
A tabletop Two Photon Absorption-Transient Current Technique (TPA-TCT) set-up built at CERN was used to investigate a non-irradiated PIN diode, an irradiated PIN diode, and a non-irradiated 5 × 5-multipad HPK LGAD. The intrinsic three dimensional spatial resolution of this method is demonstrated under normal incidence of the laser probe. A charge collection versus depth profile of the non-irradiated PIN diode is presented, where reflection on the rear silicon-air interface was observed. It is found that the time-over-threshold versus depth profile is particularly suitable to determine the boundaries of the DUT’s active volume. A depth scan of the irradiated PIN diode is discussed and a method to omit the single photon absorption background is presented. Finally, a charge collection measurement in the inter-pad region of the 5 × 5-multipad HPK LGAD is presented and it is demonstrated that TPA-TCT can be used to image the implantation and the electric field of segmented silicon devices in a three dimensional manner.
Subject
Mathematical Physics,Instrumentation
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献