Passive Hydraulic Training Simulator for Upper Arm Spasticity

Author:

Liang Jiahui1,Pei Yinan1,Ewoldt Randy H.1,Tippett Steven R.2,Hsiao-Wecksler Elizabeth T.1

Affiliation:

1. Department of Mechanical Science and Engineering, University of Illinois at Urbana-Champaign, 1206 West Green Street, Urbana, IL 61801

2. Department of Physical Therapy and Health Science, Bradley University, 1501 West Bradley Avenue, 217 Burgess Hall, Peoria, IL 61625

Abstract

Abstract Spasticity is a hypertonic muscle behavior commonly observed in patients with multiple sclerosis, cerebral palsy, stroke, etc. Clinical assessment for spasticity is done through passive stretch evaluations of various joints using qualitative clinical scales, such as the Modified Ashworth Scale (MAS). Due to the subjective nature of this evaluation method, diagnostic results can have poor reliability and inconsistency. A few research groups have developed electromechanical training simulators of upper arm spasticity with the intent of providing healthcare students practical training opportunities. This paper presents a novel, purely mechanical (nonpowered) training simulator as an alternative design approach. This passive design utilizes a hydraulic damper with selectable viscous effect to simulate the speed-dependent spastic muscle tone and a Scotch-Yoke linkage system to create the “catch-release” behavior of spasticity. An analytical fluid model was developed to systematically design the hydraulic damper. The error residuals between model prediction and experimental damping force were found within ±2.0 N and percent errors within ±10% across various testing speeds (i.e., 250, 500, 750, and 1000 mm/min). The performance of the fully assembled simulator was tested under slow (ω ≤ 60 deg/s), medium (60 deg/s < ω < 150 deg/s), and fast (ω ≥ 150 deg/s) stretch speeds, where ω is the joint angular speed. Preliminary bench-top results suggested the feasibility of replicating five distinct levels of spasticity behaviors (MAS levels 0–4), where resistive torque increased with higher stretch speed and peak resistive torque ranged from 1.3 to 6.7 N · m under the fast stretch speed.

Publisher

ASME International

Subject

Mechanical Engineering

Reference42 articles.

1. Spasticity: Clinical Perceptions, Neurological Realities and Meaningful Measurement;Pandyan;Disabil. Rehabil.,2005

2. Spasticity as an Adaptation to Pyramidal Tract Injury;Burke;Adv. Neurol.,1988

3. Upper Motor Neurone Syndrome and Spasticity

4. Measurement of Elbow Spasticity in Stroke Patients Using a Manual Spasticity Evaluator;Wu,2006

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3