Blood Damage Through a Bileaflet Mechanical Heart Valve: A Quantitative Computational Study Using a Multiscale Suspension Flow Solver

Author:

Min Yun B.1,Aidun Cyrus K.23,Yoganathan Ajit P.245

Affiliation:

1. G.W. Woodruff School of Mechanical Engineering, 801 Ferst Drive, Atlanta, GA 30332-0405 e-mail:

2. G.W. Woodruff School of Mechanical Engineering, 801 Ferst Drive, Atlanta, GA 30332-0405

3. Parker H. Petit Institute for Bioengineering and Bioscience, 315 Ferst Dr. NW, Atlanta, GA 30332 e-mail:

4. Parker H. Petit Institute for Bioengineering and Bioscience, 315 Ferst Dr. NW, Atlanta, GA 30332

5. Wallace H. Coulter Department of Biomedical Engineering, School of Chemical and Biomolecular Engineering, 313 Ferst Drive, Atlanta, GA 30332-0535 e-mail:

Abstract

Bileaflet mechanical heart valves (BMHVs) are among the most popular prostheses to replace defective native valves. However, complex flow phenomena caused by the prosthesis are thought to induce serious thromboembolic complications. This study aims at employing a novel multiscale numerical method that models realistic sized suspended platelets for assessing blood damage potential in flow through BMHVs. A previously validated lattice-Boltzmann method (LBM) is used to simulate pulsatile flow through a 23 mm St. Jude Medical (SJM) Regent™ valve in the aortic position at very high spatiotemporal resolution with the presence of thousands of suspended platelets. Platelet damage is modeled for both the systolic and diastolic phases of the cardiac cycle. No platelets exceed activation thresholds for any of the simulations. Platelet damage is determined to be particularly high for suspended elements trapped in recirculation zones, which suggests a shift of focus in blood damage studies away from instantaneous flow fields and toward high flow mixing regions. In the diastolic phase, leakage flow through the b-datum gap is shown to cause highest damage to platelets. This multiscale numerical method may be used as a generic solver for evaluating blood damage in other cardiovascular flows and devices.

Publisher

ASME International

Subject

Physiology (medical),Biomedical Engineering

Reference55 articles.

1. Mechanical Versus Bioprosthetic Valve Replacement in Middle-Aged Patients;Eur. J. Cardiothorac. Surg.,2006

2. Mechanical and Other Problems of Artificial Valves;Curr. Topic Pathol.,1994

3. Heart Valve Bioengineering,2003

4. Focused Update Incorporated Into the ACC/AHA 2006 Guidelines for the Management of Patients With Valvular Heart Disease: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the 1998 Guidelines for the Management of Patients With Valvular Heart Disease): Endorsed by the Society of Cardiovascular Anesthesiologists, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons;Circulation,2008

5. Estimation of Shear Stress-Related Blood Damage in Heart Valve Prostheses—In Vitro Comparison of 25 Aortic Valves;Int. J. Artif. Organs,1990