Affiliation:
1. Department of Clinical Chemistry, Sestre Milosrdnice University Hospital Center, Zagreb, Croatia
Abstract
Background The aim of this work was to determine the frequency of ascorbic acid-positive urine samples and to evaluate the potential risk on patient safety. Materials and methods A total of 27,856 samples submitted for urinalysis in previous year were retrospectively analysed from the laboratory information system. Urinalysis was done on Iris IQ200 analyzer using 11-parameter iChem Velocity test strips. Risk analysis combined the impact of potentially erroneous results (due to ascorbic acid interference) on patient safety (severity – S) with their frequency (occurrence – O). Potential risk was identified as high, intermediate and low. Results One thousand one hundred and ninety-nine (4.3%) and 1646 (5.9%) samples were mildly (20 mg/dL, 1+) and highly positive (40 mg/dL, 2+) for ascorbic acid, respectively. Severity (S1 to S5) was assigned to four possible errors: false-negative glucose (U-Glc: S2), blood (U-Hb: S5), nitrite (U-Nit: S4) and bilirubin (U-Bil: S1). Based on the frequency of errors, occurrence was categorized as: O2 for U-Glc, and O3 for U-Bil, U-Nit and U-Hb. The risk analysis matrix revealed that false-negative U-Hb was associated with potentially high risk, false-negative U-Nit with intermediate risk, while false-negative U-Glc and U-Bil were associated with low risk on patient safety. Conclusions Positive ascorbic acid can cause potentially high risk for patient safety by missing positive blood in urine.
Subject
Clinical Biochemistry,General Medicine
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献