Affiliation:
1. University at Buffalo, The State University of New York, USA
2. University of Plymouth, United Kingdom
3. University of Illinois at Chicago, USA
Abstract
Objective In this work, we systematically evaluated the reserved alarm sounds of the IEC 60601-1-8 international medical alarm standard to determine when and how they can be totally and partially masked. Background IEC 60601-1-8 gives engineers instruction for creating human-perceivable auditory medical alarms. This includes reserved alarm sounds: common types of alarms where each is a tonal melody. Even when this standard is honored, practitioners still fail to hear alarms, causing practitioner nonresponse and, thus, potential patient harm. Simultaneous masking, a condition where one or more alarms is imperceptible in the presence of other concurrently sounding alarms due to limitations of the human sensory system, is partially responsible for this. Methods In this research, we use automated proof techniques to determine if masking can occur in a modeled configuration of medical alarms. This allows us to determine when and how reserved alarm sound can mask other reserved alarms and to explore parameters to address discovered problems. Results We report the minimum number of other alarm sounds it takes to both totally and partially mask each of the high-, medium-, and low-priority alarm sounds from the standard. Conclusions Significant masking problems were found for both the total and partial masking of high-, medium-, and low-priority reserved alarm sounds. Application We show that discovered problems can be mitigated by setting alarm volumes to standard values based on priority level and by randomizing the timing of alarm tones.
Funder
Agency for Healthcare Research and Quality
Subject
Behavioral Neuroscience,Applied Psychology,Human Factors and Ergonomics
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献