Critical Factors of Dose Distribution in Breast Cancer Tomotherapy With Metallic Port Breast Tissue Expander: Image Correction, Delivery Mode, and Volume Impact

Author:

Lee Hsing-Yi1,Yen Yu-Hsiu23,Tsai Yu-Lun45,Tu Pei-Chih1,Pu Chi-Ming23,Lin Chia-Hong1,Lui Louis Tak14,Shaw Suzun14,Wu Ching-Jung1467,Nien Hsin-Hua348ORCID

Affiliation:

1. Department of Radiation Oncology, Oncology Treatment Center, Sijhih Cathay General Hospital, New Taipei City, Taiwan

2. Division of Plastic Surgery, Department of Surgery, Cathay General Hospital, Taipei, Taiwan

3. School of Medicine, Fu Jen Catholic University, New Taipei City, Taiwan

4. Department of Radiation Oncology, Cathay General Hospital, Taipei, Taiwan

5. Institute of Epidemiology and Preventive Medicine, College of Public Health, National Taiwan University, Taipei, Taiwan

6. Department of Radiation Oncology, National Defense Medical Center, Taipei, Taiwan

7. Department of Biomedical Engineering, I-Shou University, Kaohsiung, Taiwan

8. Institute of Biomedical Engineering, College of Electrical and Computer Engineering, National Yang Ming Chiao Tung University, Hsinchu, Taiwan

Abstract

Objectives: Breast reconstruction helps patients enhance their body image after mastectomy. Metallic ports in tissue expanders lead to dose attenuation during radiotherapy. Tissue expander volume shifts the metallic port position, possibly causing various dose alterations. This study aimed to evaluate the impact of the MAGNA-SITETM tissue expander volume on tomotherapy. Methods: Boluses and MAGNA-SITETM were placed on a Rando phantom to simulate the tissue expander under the pectoralis major. Computed tomography simulation images were transformed through replacing the electron density of (a) metallic artifact region only (Image metallic port) and (b) metallic port and artifact regions (Image Homo). Planning was calculated using fixed-beam and helical-mode techniques. Radiation was delivered with different volumes of the tissue expander. Results: Integrated 997 dose points were calculated. Planning with Image metallic port provided a calculated dose significantly closer to a realistic dose. The percentage of doses achieving the prescribed dose was significantly higher in the helical mode. In layer 2, the 100-mL tissue expander had a significantly lower measurement dose than all other volumes. Volume 150 mL had the highest increase in the measured dose difference from the plan dose at layer 2. Volume 250 mL had the highest percentage of measurement doses passing the 5% dose difference from plan dose. The coldest dose areas were noted in layers 1 and 2, especially in the metallic port–direct image mode. The average dose reduction of the measured cold areas was 6.03 ± 1.94%. Conclusion: Dose distribution was affected by the volume of the metallic port tissue expander. Tomotherapy with proper image heterogeneity correction and helical mode can reduce the attenuation from the metallic port. A tissue expander volume of 150 to 250 mL is suitable. Patients with high risk at the chest wall should be evaluated carefully to avoid underdosing. Radiation oncologists should closely cooperate with plastic surgeons to optimize treatment for each patient.

Publisher

SAGE Publications

Subject

Cancer Research,Oncology

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3