Algorithm for Working with Metal Objects in Breast Expander in Radiotherapy

Author:

Panchenko V. R.1,Lebedeva Zh. S.2,Litvinov A. P.2

Affiliation:

1. Peter the Great St.Petersburg Polytechnic University

2. Pavlov First Saint Petersburg State Medical University

Abstract

Purpose: The article is devoted to the analysis of the developed algorithm of contouring minimizing dosimetric deviations during post-mastectomy radiation therapy of patients with breast tissue expanders with a metal port. The influence of the main errors in contouring the critical structures of the above patients on the quality of dosimetric planning is also considered. Material and methods: Mentor CPX4 tissue expander with a two component SmoothCast medical plastic shell imitating the patient's skin was scanned on three protocols using CT GE Optima 540, the resulting series of slices were contoured by different methods in the Eclipse planning systemand dosimetry evaluated by film dosimetry. Algorithm minimizing a contouring errors was developed. Results: When contouring on the basis of the developed algorithm, the deviation in dose compared with the value calculated in the planning system was 1.7 %. Contouring on the basis of the above errors gave a deviation in the dose up to 6.7 %. Conclusion: The developed delineation algorithm can be used both in departments with the presence of CT with the “extended curve” function, and in departments where this function is unavailable in CT.

Publisher

Association of Medical Physicists in Russia

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3