Early shape change behaviour of an uncemented contemporary hip cup: A cadaveric experiment replicating host bone behaviour through temperature control

Author:

Manning William A1,Pandorf Thomas2,Deehan David J1,Holland James1

Affiliation:

1. Newcastle Surgical Training Centre, Freeman Hospital, Newcastle upon Tyne, UK

2. Medical Products Division, CeramTec GmbH, Plochingen, Germany

Abstract

Modular uncemented acetabular components are in common use. Fixation is dependent upon press-fit but the forces necessary to achieve initial stability of the construct at implantation may deform the shell and prevent optimal seating of the polyethylene liner insert. Previous work using single-time point measurements in uncontrolled ambient temperature poorly replicates the native state. A controlled study was performed to observe the time-dependent behaviour of an uncemented acetabular shell in the early phase after implantation into the human acetabulum at near physiological temperature. Using a previously validated cadaveric hip model at controlled near physiological temperature with standardised surgical technique, immediate and delayed shell geometry was determined. Eight custom made 3-mm-thick titanium alloy (TiAl6V4) shells were implanted into four cadavers (eight hips). Time-dependent shell deformation was determined using the previously validated ATOS Triple Scan III (ATOS) optical measurement system. The pattern of change in the shape of the surgically implanted shell was measured at three time points after insertion. We found a consistent pattern for quantitative and directional deformation of the shells. In addition, there was consistency for relaxation of the deformation with time. Immediate mean change in shell radius was 104 µm (standard deviation 32, range 67–153) relaxing to mean 96 µm (standard deviation 32, range 63–150) after 10 min and mean 92 µm (standard deviation 28, range 66–138) after 20 min. The clinical significance of this work is the finding of a time-dependent early deformation of acetabular titanium shells on insertion adjusted for near physiological temperature-controlled host bone.

Funder

ceramtec GmBH

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3