Affiliation:
1. Research Dept. Diasonics, Inc. Milpitas, CA 95035
Abstract
In this paper, a theoretical framework is developed which removes the Gaussian assumption commonly used in ultrasonic attenuation estimation based on mean frequency shift. The theory is developed for general non-Gaussian spectra and for media whose attenuation coefficient is nonlinear with frequency. Then, a linear approximation in the estimation of the attenuation coefficient's slope is examined. It is shown that the error due to the linear approximation is negligibly small.
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献