Affiliation:
1. Department of Radiology The University of Texas Medical School 6431 Fannin Street Houston, TX 77030
Abstract
The relationship between center frequency downshift and transmitted bandwidth was investigated for a pulse with Gaussian amplitude spectrum propagating through a lossy medium with power law frequency dependence of attenuation. Tissue equivalent material was characterized by multiple narrowband attenuation measurements, via a substitution method. Power law curves were fitted to the data. The parameters of the curves were used to predict the behavior of the center frequency downshift vs. transmitted bandwidth obtained from a second experiment. The results confirm the mathematical model.
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Reference9 articles.
1. Ultrasonic Attenuation Tomography of Soft Tissues
2. Acoustic properties of normal and cancerous human liver—I. Dependence on pathological condition
3. Jones J. P., Beherens M., Ferrari L., Gonzales V. In Vivo Differentiation of Abdominal Organs and the Early Detection of Cirrhosis of the Liver by the Quantitative Analysis of A-mode Ultrasound Waveforms, in Proc. 26th Ann. Conv. AIUM, p. 68, August, 1981 (Abstract only).
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献