Elimination of Microbubbles from the Extracorporeal Circuit: Dynamic Bubble Trap versus Arterial Filter

Author:

Martens S.1,Dietrich M.1,Pietrzyk R.2,Graubitz K.1,Keller H.1,Moritz A.1

Affiliation:

1. Department for Thoracic and Cardiovascular Surgery, University Hospital J.W. Goethe, Frankfurt am Main - Germany

2. HP-Medica Inc., Augsburg - Germany

Abstract

Background Open heart surgery is associated with important risk of cerebral and peripheral organ dysfunction, attributed in part to microbubbles generated in or not eliminated from the ECC. For elimination of microbubbles, a dynamic bubble trap (DBT(tm)) was developed for the arterial line of ECCs. Methods Bubble eliminating properties of an arterial filter were evaluated in four CABG patients and compared to the performance of the DBT in four patients. One patient received both devices. Results Elimination of bubbles between 40–120μm was significantly higher with the DBT (88% vs. 57% with arterial filter, p=0.034). Reduction of bubbles below 40μm was equivalent in both groups. The combination of both devices was most effective (94% for bubbles >40μm). Conclusion Arterial filter and DBT are equally effective in elimination of smaller gas bubbles. However, bigger bubbles possibly causing cerebral and peripheral organ damage are eliminated to a greater degree by the DBT.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering

Cited by 7 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3