Affiliation:
1. Kerckhoff-Klinik, Thoracic and Cardiovascular Surgery, Bad Nauheim
2. Convergenza AG, Vaduz
Abstract
Air microbubbles mostly occur unnoticed during cardiopulmonary bypass and are predominantly responsible for serious postoperative psycho- neurological dysfunction. A dynamic bubble trap (DBT), which removes air microbubbles from the arterial blood, was tested in a clinical study. The aim was to evaluate the efficiency of microbubble removal under clinical conditions. As blood passes through the DBT, which is placed in the arterial line between the arterial filter and arterial cannula, it is converted into a rotating stream. The bubbles are directed to the centre of the blood flow and are collected in the distal end of the DBT, from where they are returned to the cardiotomy reservoir. Doppler ultrasonography was used to detect the microbubbles before and after the DBT, and also the number of high- intensity transient signals (HITS) in the right and left middle cerebral artery during extracorporeal circulation. A significant reduction of microbubbles in the arterial line (3990 before DBT, 537 after, p < 0.001) and HITS in the brain (51 in the DBT group, 77 in the placebo group, p = 0.04) was measured.
Subject
Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine
Cited by
17 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献