Characterization of a MLIC Detector for QA in Scanned Proton and Carbon Ion Beams

Author:

Vai Alessandro1,Mirandola Alfredo1,Magro Giuseppe1,Maestri Davide1,Mastella Edoardo1,Mairani Andrea12,Molinelli Silvia1,Russo Stefania1,Togno Michele34,Civita Sara La35,Ciocca Mario1

Affiliation:

1. Fondazione CNAO (Italian National Center for Oncological Hadronterapy), Pavia, Italy

2. Heidelberg Ion-Beam Therapy Center, Heidelberg, Germany

3. R&D Department, IBA Dosimetry, Schwarzenbruck, Germany

4. Now at Center for Proton Therapy, Paul Scherrer Institut, Villigen, Switzerland

5. Now at Mevion Medical System BV, Maastricht, Holland

Abstract

AbstractPurpose:Beam energy validation is a fundamental aspect of the routine quality assurance (QA) protocol of a particle therapy facility. A multilayer ionization chamber (MLIC) detector provides the optimal tradeoff between achieving accuracy in particle range determination and saving operational time in measurements and analysis procedures. We propose the characterization of a commercial MLIC as a suitable QA tool for a clinical environment with proton and carbon-ion scanning beams.Materials and Methods:Commercial MLIC Giraffe (IBA Dosimetry, Schwarzenbruck, Germany) was primarily evaluated in terms of short-term and long-term stability, linearity with dose, and dose-rate independence. Accuracy was tested by analyzing range of integrated depth-dose curves for a set of representative energies against reference acquisitions in water for proton and carbon ion beams; in addition, 2 modulated proton spread-out Bragg peaks were also measured. Possible methods to increase the native spatial resolution of the detector were also investigated.Results:Measurements showed a high repeatability: mean relative standard deviation was within 0.5% for all channels and both particle types. The long-term stability of the gain calibration showed discrepancies less than 1% at different times. The detector response was linear with dose (R2 > 0.99) and independent on the dose rate. Measurements of integrated depth-dose curve ranges revealed a mean deviation from reference measurements in water of 0.1 ± 0.3 mm for protons with a maximum difference of 0.4 mm and 0.2 ± 0.6 mm with maximum difference of 0.85 mm for carbon ion beams. For the 2 modulated proton spread-out Bragg peaks, measured differences in distal dose falloff were ≤0.5 mm against calculated values.Conclusions:The detector is stable, linearly responding with dose, precise, and easy to handle for QA beam energy checks of proton and carbon ion beams.

Publisher

International Journal of Particle Therapy

Subject

Radiology Nuclear Medicine and imaging,Atomic and Molecular Physics, and Optics

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3