Detection of fast neutrons from D–T nuclear reaction using a 4H–SiC radiation detector

Author:

Zatko Bohumir1,Sagatova Andrea23,Sedlackova Katarina2,Necas Vladimir2,Dubecky Frantisek1,Solar Michael4,Granja Carlos4

Affiliation:

1. Institute of Electrical Engineering, Slovak Academy of Sciences, Dúbravská cesta 9, SK-841 04 Bratislava, Slovak Republic

2. Institute of Nuclear and Physical Engineering, Faculty of Electrical Engineering and Information Technology, Slovak University of Technology in Bratislava, Ilkovičova 3, SK-81219 Bratislava, Slovak Republic

3. University Centre of Electron Accelerators, Slovak Medical University, Ku kyselke 497, SK-911 06 Trenčín, Slovak Republic

4. Institute of Experimental and Applied Physics, Czech Technical University in Prague, Horská 3a/22, CZ -128 00 Prague 2, Czech Republic

Abstract

The particle detector based on a high purity epitaxial layer of 4H–SiC exhibits promising properties in detection of various types of ionizing radiation. Due to the wide band gap of 4H–SiC semiconductor material, the detector can reliably operate at room and also elevated temperatures. In this work we focused on detection of fast neutrons generated the by D–T (deuterium–tritium) nuclear reaction. The epitaxial layer with a thickness of 105 [Formula: see text]m was used as a detection part. A circular Schottky contact of a Au/Ni double layer was evaporated on both sides of the detector material. The detector structure was characterized by current-voltage and capacitance-voltage measurements, at first. The results show very low current density (<0.1 nA/cm[Formula: see text] at room temperature and good homogeneity of free carrier concentration in the investigated depth. The fabricated detectors were tested for detection of fast neutrons generated by the D–T reaction. The energies of detected fast neutrons varied from 16.0 MeV to 18.3 MeV according to the acceleration potential of deuterons, which increased from 600 kV up to 2 MV. Detection of fast neutrons in the SiC detector is caused by the elastic and inelastic scattering on the silicon or carbide component of the detector material. Another possibility that increases the detection efficiency is the use of a conversion layer. In our measurements, we glued a HDPE (high density polyethylene) conversion layer on the detector Schottky contact to transform fast neutrons to protons. Hydrogen atoms contained in the conversion layer have a high probability of interaction with neutrons through elastic scattering. Secondary generated protons flying to the detector can be easily detected. The detection properties of detectors with and without the HDPE conversion layer were compared.

Publisher

World Scientific Pub Co Pte Lt

Cited by 10 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献