Affiliation:
1. Individual Monitoring Service, Helmholtz Zentrum München – German Research Center for Environmental Health (GmbH), Otto-Hahn-Ring 6, 81739 Münich, Germany
Abstract
Abstract
Radiation qualities were characterised and validated at the fully automated X-ray calibration facility of the Individual Monitoring Service at Helmholtz Zentrum München by using half-value layer and Hp(10) dosimetry approaches specified in the updated ISO 4037:2019 standard. As the ISO 4037 contains a somewhat vague description of the half-value layer procedure, we extended it to be more constrained and thus less subjective in its implementation. We specify both the measurement and data analysis steps performed in order to provide reproducible half-value layer results and compare the results with the Hp(10) dosimetry-based validation approach specified in the ISO 4037 as an alternative for field validation. Finally, we discuss the implications of our results and the extended procedure on validation and the recent changes to the ISO 4037.
Publisher
Oxford University Press (OUP)
Subject
Public Health, Environmental and Occupational Health,Radiology Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology
Cited by