Affiliation:
1. Individual Monitoring Service, Helmholtz Zentrum München – German Research Center for Environmental Health (GmbH), Otto-Hahn-Ring 6, München, Germany
Abstract
Abstract
The individual monitoring service at the Helmholtz Zentrum München has extended the calibration facilities provided at the Munich IAEA/WHO secondary standard dosimetry laboratory with a new fully automated X-ray calibration facility for the calibration of personal dosemeters according to the ISO 4037 standard series. This work describes the X-ray irradiation system and the automated dosemeter transport system as well as the measurements performed to characterize the photon fields and to ascertain conformity to the ISO standard. Standard uncertainties of the radiation quantities provided by the system are given together with a discussion of some of the problems that are encountered in fulfilling the requirements of a matched field according to the latest drafts of the ISO 4037-1 standard.
Publisher
Oxford University Press (OUP)
Subject
Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献