Abstract
Abstract
To provide a data base for the establishment of a method to detect potentially incorrect SMA 6 test results, we analyzed 2970 pairs of SMA 6 results. For each test result, as well as for the urea nitrogen/creatinine ratio and the anion gap, we estimated to probability density function of the amount of change since the most recent test of the same type. We made separate estimates for each of three ranges of the current test result and two ranges of time-spacing between the tests. Tables of nonparametric estimates of selected points from each of the six distributions for each test result and calculated quantity were developed. These tables could be used to identify unlikely test results, either manually or with a laboratory computer.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
18 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. "Delta check" values and its clinic and laboratory use;Zbornik radova 26. medunarodni kongres Mediteranske federacije za zdravlje i produkciju preživara - FeMeSPRum - zbornik radova;2024
2. Estimación de los límites de cambio (deltacheck) en el laboratorio clínico;Advances in Laboratory Medicine / Avances en Medicina de Laboratorio;2021-07-21
3. Estimation of change limits (deltacheck) in clinical laboratory;Advances in Laboratory Medicine / Avances en Medicina de Laboratorio;2021-04-01
4. Evidence-based approach to setting delta check rules;Critical Reviews in Clinical Laboratory Sciences;2020-08-14
5. Delta Checks in the clinical laboratory;Critical Reviews in Clinical Laboratory Sciences;2019-01-11