Abstract

Abstract A case is presented of a patient with stage D prostatic carcinoma, from whom a serum sample proved to be an outlier in a correlation study performed with a 2nd-generation prostate-specific antigen (PSA) assay on the Immulite system (6.4 micrograms/L) and IMx (101 micrograms/L). Clearly, the PSA result reported by Immulite was falsely low. For nine longitudinal samples, Immulite results were approximately 20-fold lower than the IMx value (range of IMx results 5-275 micrograms/L). A selection of the samples was analyzed with the ACS:180, ES-600, and IMx (all > 180 micrograms/L); Immulite, DPC Coat-A-Count IRMA, Immuno 1, AIA-pack, and Tandem-R (all <70 micrograms/L); and Immulite free PSA assay (41 micrograms/L). Gel filtration demonstrated that apart from the alpha1-antichymotrypsin (ACT) complex, no other complexes were found. However, the sample consisted of 53% free PSA (IMx). Possibly, a change of conformation of the PSA molecule resulted in a decreased binding to ACT and a reduced affinity of the antibodies used in the affected assays.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3