Validation of Linear Range HER2/Estrogen Receptor/Progesterone Receptor IHControls for Daily Quality Assurance

Author:

Sompuram Seshi R1,Vani Kodela1,Ryan Lori2,Johnson Corissa2,Szabolcs Matthias3,Peruyero Leonore3,Balaton André4,Pierrot Sandrine4,Joseph Lija5,Pilichowska Monika6,Naber Stephen6,Goldsmith Jeffrey7,Green Samantha8,Bogen Steve A1

Affiliation:

1. Boston Cell Standards , Boston MA , USA

2. Allina Health , Minneapolis, MN , USA

3. Columbia University Medical Center , New York, NY , USA

4. Praxea Diagnostics , Massy , France

5. Lowell General Hospital , Lowell, MA , USA

6. Tufts Medical Center , Boston, MA , USA

7. Boston Children’s Hospital , Boston, MA , USA

8. Catholic Medical Center , Manchester, NH , USA

Abstract

AbstractObjectivesTo evaluate a new US Food and Drug Administration (FDA)–cleared immunohistochemistry (IHC) control (IHControls [Boston Cell Standards]) comprising peptide epitopes for HER2, estrogen receptor (ER), and progesterone receptor (PR) attached to cell-sized microspheres and to compare its performance against conventional tissue controls.MethodsIHControls and tissue/cell line controls for HER2, ER, and PR were compared side by side daily at 5 clinical IHC laboratories for 1 to 2 months. Separately, the sensitivity of the 2 types of controls was evaluated in simulated IHC assay failure experiments by diluting the primary antibody. Additional evaluations included lot-to-lot manufacturing reproducibility of 3 independent lots and specificity against 26 antigenically irrelevant IHC stains.ResultsSide-by-side testing revealed a 99.6% concordance between IHControls and tissue controls across 5 IHC laboratories and 766 individual evaluations. Three discordant quality control events were the result of operator error. Simulated assay failure data showed that both IHControls and tissue controls are similarly capable of detecting IHC staining errors. Manufacturing reproducibility of IHControls showed less than 10% variability (coefficient of variation). No cross-reactions were detected from 26 antigenically irrelevant IHC stains.ConclusionsIHControls, the first FDA-cleared IHC controls, can sensitively and accurately detect IHC assay problems, similar to tissue controls.

Funder

National Cancer Institute

National Institutes of Health

Publisher

Oxford University Press (OUP)

Subject

General Medicine

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3