Extraction outcomes of implantable cardioverter-defibrillator leads vary by manufacturer and model family

Author:

Hayashi Katsuhide1ORCID,Callahan Thomas1ORCID,Rickard John1,Younis Arwa1ORCID,Baranowski Bryan1,Martin David1,Nakhla Shady1ORCID,Tabaja Chadi1,Wilkoff Bruce L1ORCID

Affiliation:

1. Cardiac Electrophysiology and Pacing Section, Department of Cardiovascular Medicine, Cleveland Clinic , 9500 Euclid Avenue Desk J2-2, Cleveland, OH 44195 , USA

Abstract

Abstract Aims Transvenous lead extraction (TLE) of implantable cardioverter-defibrillator (ICD) leads is considered challenging. The structure of each ICD leads is variable between manufacturer and model families. The net impact of lead family on the safety and effectiveness of TLE is poorly characterized. We assessed the safety and efficacy of ICD TLE and the impact of manufacturer ICD model family on the outcomes. Methods and results The study cohort included all consecutive patients with ICD who underwent TLE between 2013 and 2022 and are enrolled in the Cleveland Clinic Prospective TLE Registry. A total of 885 ICD leads (median implant duration 8 years) in 810 patients were included. Complete ICD TLE success was achieved in 97.2% of the leads (n = 860) and in 98.0% of the patients (n = 794). Major complications occurred in 22 patients (2.7%). Complete procedural success rate varied by manufacturer and lead family; Medtronic 98.9%, Abbott 95.9%, Boston Scientific 95.0%, Biotronik 91.2%, P = 0.03, and Linox family leads had the lowest, 89.7% P = 0.02. Multivariable predictors of incomplete ICD lead removal included ICD lead age > 10 years and Linox family lead. Multivariable predictors of major complications included ICD lead age > 15 years and longer lead extraction time, and predictors of all-cause mortality within 30 days included lead extraction for infection, end-stage renal disease, and higher New York Heart Association functional class. Conclusion Complete and safe ICD lead removal rate by TLE is extremely high but varied by manufacturer and lead family. Linox family lead and >10 years lead age were independent predictors of incomplete lead removal.

Funder

Deb family endowed chair in Lead Management

Publisher

Oxford University Press (OUP)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3