RADIATION DOSE OF THE EYE LENS IN CT EXAMINATIONS OF THE BRAIN IN CLINICAL PRACTICE—THE EFFECT OF RADIOGRAPHER TRAINING TO OPTIMISE GANTRY TILT AND SCAN LENGTH

Author:

Tarkiainen Jeremias1ORCID,Nadhum Miia12,Heikkilä Annele12,Rinta-Kiikka Irina3,Joutsen Atte2ORCID

Affiliation:

1. Faculty of Medicine and Health Technology, Tampere University , Tampere , Finland

2. Department of Medical Physics, Medical Imaging Center, Pirkanmaa Hospital District , Tampere , Finland

3. Department of Radiology, Medical Imaging Center, Pirkanmaa Hospital District , Tampere , Finland

Abstract

AbstractLenses are always exposed to radiation in brain computed tomography (CT) scans. However, the lens dose can be reduced by excluding lens from scanning area by optimising gantry tilt and scan length. The object of this study is to retrospectively analyse if the optimisation by gantry tilt and scan length have been adequate in the CT scan of the brain, and to prospectively analyse the effect of radiographer training to the quality of the CT examinations. This study was conducted in two parts. In all, 329 brain CTs performed in the Tampere University Hospital from 2017 to 2019 were revised retrospectively. The prospective part included 51 brain CT studies conducted in October 2021. Dose to the eye of the lens was modelled using CT-Expo using zero-degree beam angle and scan lengths to expose the lens either to the primary or scattered radiation. Non-zero gantry tilt had been used in a large proportion of the CT examinations in the retrospective setting, 84.8%. However, the lenses were successfully excluded from the scan area in only 1.8% of the examinations. In the prospective part, the gantry tilt was used in 98% of the studies and the proportion of successful examinations rose from 1.8 to 11.8%. The lens dose decreased significantly when the eyes were excluded from the imaging area. The modelled lens dose in the large retrospective part was 25.9 mGy (17.8–49.2 mGy) when the eyes were included and 1.5 mGy (0.4–1.9 mGy) when the eyes were excluded. The lens dose was similar in the small prospective part. Despite the gantry tilt is widely used, unnecessary lens irradiation occurs extensively because of suboptimal gantry tilt and scan length. The training of radiographers reduces the radiation exposure to the lens by more optimal gantry tilt and scan length.

Funder

Tampere University Hospital

Publisher

Oxford University Press (OUP)

Subject

Public Health, Environmental and Occupational Health,Radiology, Nuclear Medicine and imaging,General Medicine,Radiation,Radiological and Ultrasound Technology

Reference34 articles.

1. European trends in radiology: investigating factors affecting the number of examinations and the effective dose;Masjedi;Radiol. Med.,2020

2. European Commission (EC). Medical radiation exposure of the European population;Takase;Radi. Prot. 180. (2014),2010

3. Trends of computed tomography use among children in Finland;Abuhamed;Eur. J. Radiol. Open,2020

4. The population effective dose of medical computed tomography examinations in Taiwan for 2013;Yeh;PLoS One,2016

5. ICRP publication 118: ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs–threshold doses for tissue reactions in a radiation protection context;Clement;Ann. ICRP,2012

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3