Objective. Arterial oxygen saturation (Sao2) and arterial carbon dioxide partial pressure (Paco2) are 2 of the most important respiratory parameters in the treatment of critically ill neonates. Noninvasive monitoring of these parameters is desirable for continuous estimating of the respiratory status and reducing blood loss because of repeated blood gas analyses. Transcutaneous measurement of Pco2 (Ptcco2) represents a simple and noninvasive technique for continuous monitoring of ventilation. However, sensor preparation, positioning, taping, and repeated changes of the sensor location make the handling difficult and complicate its use in the neonatal care unit. Recently, a new sensor for combined assessment of pulse oximetry oxygen saturation (Spo2) and Ptcco2 has been introduced (TOSCA Monitor; Linde Medical Sensors, Basel, Switzerland). The monitor combines pulse oximetry and Ptcco2 measurement in a single ear sensor, which works at 42°C to enhance blood flow in capillaries below the sensor.
Methods. In a prospective, open, nonrandomized study of 60 ill neonates, the new ear sensor for combined assessment of Spo2 and Ptcco2 at 42°C was tested. The sensor was adapted to the ear of a neonate with a Varihesive layer (Conva Tec; Princeton, NJ). Data obtained from the ear sensor were compared with Spo2 Finger/Heel, Sao2, and Paco2 obtained from arterial blood gas in 30 patients and with a capillary blood gas in an additional 30 patients using Bland Altman bias analysis. Data are presented as median (range).
Results. The postconceptional age of the patients was 38.3 weeks (range: 28 5/7–40 5/7) in the arterial group and 37.9 weeks (range: 29 6/7–41 0/7) in the capillary group. Age of the newborns studied was 3.5 days (range: 1–28) in the arterial blood sample group (n = 30) and 6 days (range: 2–28) in the capillary blood sample group (n = 30). Patient weight was 3.02 kg (range: 1.5–4.5) in the arterial group and 2.76 kg (range: 1.0–3.71) in the other group. Three patients had weights of <1500 g. Twenty-one of 60 patients were conventionally ventilated, 4 patients received high-frequency oscillation, and 35 were not ventilated. Mean difference (bias) and precision (2 SD of the mean difference) between Ptcco2 TOSCA and Paco2 were −0.44 kPa (−3.21 mm Hg) and 0.82 kPa (6.02 mm Hg) and between Ptcco2 TOSCA and Pcapco2 were −0.09 kPa (−0.67 mm Hg) and 1.11 kPa (8.07 mm Hg), respectively. Spo2 assessment by the TOSCA revealed slightly higher values compared with Sao2 (bias: −0.48%), whereas Spo2 Finger/Heel values were slightly lower than Sao2 (bias: 0.52%).
Conclusion. The TOSCA monitor with the ear sensor adapted to ears of neonates allows reliable estimation of Sao2 and Paco2. A potential benefit is the reduction in motion artifacts because of less head movement in newborns and that only a single cable leads form the patient to the monitor. In addition, the sensor is not removed for chest radiograph or for nursing the infant on his or her parent's lap. Long-term studies in a large population with continuous measurements are required to confirm these preliminary findings and to elucidate the benefits in detection of respiratory deterioration and the potential side effects of this sensor.
Publisher
American Academy of Pediatrics (AAP)
Subject
Pediatrics, Perinatology and Child Health
Reference14 articles.
1. Bucher HU, Fanconi S, Baeckert P, et al. Hyperoxemia in newborn infants: detection by pulse oximetry. Pediatrics. 1989;84:226–230
2. Bucher HU, Fanconi S, Fallenstein F, et al. Transcutaneous carbon dioxide tension in newborn infants: reliability and safety of continuous 24-hour measurement at 42°C. Pediatrics. 1986;78:631–635
3. Carter B, Hochmann M, Osborne A, et al. A comparison of two transcutaneous monitors for the measurement of arterial PO2 and PCO2 in neonates. Anaesth Intensive Care. 1995;23:708–714
4. Eberhard P, Gisiger PA, Gardaz JP, et al. Combining transcutaneous blood gas measurement and pulse oximetry. Anesth Analg. 2002;94:S76–S80
5. Dullenkopf A, Di Bernardo S, Berger F, et al. Evaluation of a new combined SpO2/PtcCO2 sensor in anaesthetized paediatric patients. Paediatr Anaesth. 2003;13:1–8
Cited by
34 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献