Author:
Chang Ruth P,Smith Anne,Kent Peter,Saraceni Nic,Hancock Mark,O’Sullivan Peter B,Campbell Amity
Abstract
Abstract
Background
Wearable sensor technology may allow accurate monitoring of spine movement outside a clinical setting. The concurrent validity of wearable sensors during multiplane tasks, such as lifting, is unknown. This study assessed DorsaVi Version 6 sensors for their concurrent validity with the Vicon motion analysis system for measuring lumbar flexion during lifting.
Methods
Twelve participants (nine with, and three without back pain) wore sensors on T12 and S2 spinal levels with Vicon surface markers attached to those sensors. Participants performed 5 symmetrical (lifting from front) and 20 asymmetrical lifts (alternate lifting from left and right). The global-T12-angle, global-S2-angle and the angle between these two sensors (relative-lumbar-angle) were output in the sagittal plane. Agreement between systems was determined through-range and at peak flexion, using multilevel mixed-effects regression models to calculate root mean square errors and standard deviation. Mean differences and limits of agreement for peak flexion were calculated using the Bland Altman method.
Results
For through-range measures of symmetrical lifts, root mean squared errors (standard deviation) were 0.86° (0.78) at global-T12-angle, 0.90° (0.84) at global-S2-angle and 1.34° (1.25) at relative-lumbar-angle. For through-range measures of asymmetrical lifts, root mean squared errors (standard deviation) were 1.84° (1.58) at global-T12-angle, 1.90° (1.65) at global-S2-angle and 1.70° (1.54) at relative-lumbar-angle. The mean difference (95% limit of agreement) for peak flexion of symmetrical lifts, was − 0.90° (-6.80 to 5.00) for global-T12-angle, 0.60° (-2.16 to 3.36) for global-S2-angle and − 1.20° (-8.06 to 5.67) for relative-lumbar-angle. The mean difference (95% limit of agreement) for peak flexion of asymmetrical lifts was − 1.59° (-8.66 to 5.48) for global-T12-angle, -0.60° (-7.00 to 5.79) for global-S2-angle and − 0.84° (-8.55 to 6.88) for relative-lumbar-angle.
Conclusion
The root means squared errors were slightly better for symmetrical lifts than they were for asymmetrical lifts. Mean differences and 95% limits of agreement showed variability across lift types. However, the root mean squared errors for all lifts were better than previous research and below clinically acceptable thresholds. This research supports the use of lumbar flexion measurements from these inertial measurement units in populations with low back pain, where multi-plane lifting movements are assessed.
Publisher
Springer Science and Business Media LLC
Subject
Orthopedics and Sports Medicine,Rheumatology
Reference39 articles.
1. Wu A, March L, Zheng X, Huang J, Wang X, Zhao J, et al. Global low back pain prevalence and years lived with disability from 1990 to 2017: estimates from the Global Burden of Disease Study 2017. Ann Transl Med. 2020;8(6):299.
2. Hartvigsen J, Hancock MJ, Kongsted A, Louw Q, Ferreira ML, Genevay S, et al. What low back pain is and why we need to pay attention. Lancet. 2018;391(10137):2356–67.
3. Christe G, Aussems C, Jolles BM, Favre J. Patients with chronic low back pain have an individual movement signature: A comparison of angular amplitude, angular velocity and muscle activity across multiple functional tasks. Front Bioeng Biotechnol. 2021;9(1109):13.
4. Laird RA, Keating JL, Kent P. Subgroups of lumbo-pelvic flexion kinematics are present in people with and without persistent low back pain. BMC Musculoskelet Disord. 2018;19(1):309.
5. Laird R, Gilbert J, Kent P, Keating J. Comparing lumbo-pelvic kinematics in people with and without back pain: A systematic review and meta-analysis. BMC Musculoskelet Disord. 2014;15:229.
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献