Author:
Chin Robin,Hsiao-Wecksler Elizabeth T,Loth Eric,Kogler Géza,Manwaring Scott D,Tyson Serena N,Shorter K Alex,Gilmer Joel N
Abstract
Abstract
Background
A self-contained, self-controlled, pneumatic power harvesting ankle-foot orthosis (PhAFO) to manage foot-drop was developed and tested. Foot-drop is due to a disruption of the motor control pathway and may occur in numerous pathologies such as stroke, spinal cord injury, multiple sclerosis, and cerebral palsy. The objectives for the prototype PhAFO are to provide toe clearance during swing, permit free ankle motion during stance, and harvest the needed power with an underfoot bellow pump pressurized during the stance phase of walking.
Methods
The PhAFO was constructed from a two-part (tibia and foot) carbon composite structure with an articulating ankle joint. Ankle motion control was accomplished through a cam-follower locking mechanism actuated via a pneumatic circuit connected to the bellow pump and embedded in the foam sole. Biomechanical performance of the prototype orthosis was assessed during multiple trials of treadmill walking of an able-bodied control subject (n = 1). Motion capture and pressure measurements were used to investigate the effect of the PhAFO on lower limb joint behavior and the capacity of the bellow pump to repeatedly generate the required pneumatic pressure for toe clearance.
Results
Toe clearance during swing was successfully achieved during all trials; average clearance 44 ± 5 mm. Free ankle motion was observed during stance and plantarflexion was blocked during swing. In addition, the bellow component repeatedly generated an average of 169 kPa per step of pressure during ten minutes of walking.
Conclusion
This study demonstrated that fluid power could be harvested with a pneumatic circuit built into an AFO, and used to operate an actuated cam-lock mechanism that controls ankle-foot motion at specific periods of the gait cycle.
Publisher
Springer Science and Business Media LLC
Subject
Health Informatics,Rehabilitation
Reference21 articles.
1. Perry J: Gait analysis: normal and pathological function. Thorofare, NJ: Slack; 1992.
2. Newman CJ, Walsh M, O'Sullivan R, Jenkinson A, Bennett D, Lynch B, O'Brien T: The characteristics of gait in Charcot-Marie-Tooth disease types I and II. Gait Posture 2007, 26: 120-127.
3. American Academy of Orthopaedic Surgeons: Atlas of orthoses and assistive devices. 3rd edition. St. Louis: Mosby; 1997.
4. Lusardi MM, Nielsen CC: Orthotics and prosthetics in rehabilitation. Boston: Butterworth-Heinemann; 2000.
5. Yamamoto S, Hagiwara A, Mizobe T, Yokoyama O, Yasui T: Development of an ankle – foot orthosis with an oil damper. Prosthet Orthot Int 2005, 29: 209-219.
Cited by
57 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献