Novel methodology to perform sulfur hexafluoride (SF6)-based multiple-breath wash-in and washout in infants using current commercially available equipment

Author:

Gustafsson P. M.12,Robinson P. D.34,Lindblad A.25,Oberli D.6

Affiliation:

1. Department of Pediatrics, Central Hospital, Skövde, Sweden;

2. The Sahlgrenska Academy at the University of Gothenburg, Gothenburg, Sweden;

3. Department of Respiratory Medicine, The Children's Hospital at Westmead, Sydney, New South Wales, Australia;

4. Discipline of Paediatrics and Child Health, Sydney Medical School, University of Sydney, Australia;

5. CF Centre, Queen Silvia Children's Hospital, Gothenburg, Sweden; and

6. ECO MEDICS AG, Duernten, Switzerland

Abstract

Multiple-breath inert gas washout (MBW) is ideally suited for early detection and monitoring of serious lung disease, such as cystic fibrosis, in infants and young children. Validated commercial options for the MBW technique are limited, and suitability of nitrogen (N2)-based MBW is of concern given the detrimental effect of exposure to pure O2 on infant breathing pattern. We propose novel methodology using commercially available N2 MBW equipment to facilitate 4% sulfur hexafluoride (SF6) multiple-breath inert gas wash-in and washout suitable for the infant age range. CO2, O2, and sidestream molar mass sensor signals were used to accurately calculate SF6 concentrations. An improved dynamic method for synchronization of gas and respiratory flow was developed to take into account variations in sidestream sample flow during MBW measurement. In vitro validation of triplicate functional residual capacity (FRC) assessments was undertaken under dry ambient conditions using lung models ranging from 90 to 267 ml, with tidal volumes of 28-79 ml, and respiratory rates 20–60 per minute. The relative mean (SD, 95% confidence interval) error of triplicate FRC determinations by washout was −0.26 (1.84, −3.86 to +3.35)% and by wash-in was 0.57 (2.66, −4.66 to +5.79)%. The standard deviations [mean (SD)] of percentage error among FRC triplicates were 1.40 (1.14) and 1.38 (1.32) for washout and wash-in, respectively. The novel methodology presented achieved FRC accuracy as outlined by current MBW consensus recommendations (95% of measurements within 5% accuracy). Further clinical evaluation is required, but this new technique, using existing commercially available equipment, has exciting potential for research and clinical use.

Funder

2015 Vertex Innovation Grant - All authors except Daniel Oberli

Publisher

American Physiological Society

Subject

Physiology (medical),Physiology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3