Affiliation:
1. Department of Medical Physics , Maria Sklodowska-Curie National Research Institute of Oncology (MSCNRIO) in Warsaw , Poland
Abstract
Abstract
Introduction: Dosimetric verification of Intensity Modulation Radiotherapy treatment plans is usually carried out before the start of treatment. It is of special importance in the case of highly modulated plans, such as plans for patients in whom many tumors are irradiated simultaneously. In this work, we present the results of the verification of multi-target plans performed with the stereotactic HyperArc technique.
Material and methods: The results of dosimetric verification of 35 patient plans in the head are presented. The results are analyzed in terms of the number of tumors, and the distance of the tumor from the isocenter. Measurements were carried out with the film method. The gamma methodology was used (3%/2mm).
Results: The results showed a very good agreement between measurements and calculations.
Conclusions: No dependence of the verification result on the number of targets and the distance between the center of tumor and isocenter was found.
Reference13 articles.
1. 1. Shah AP, Meeks DT, Willoughby TR, et al. Intrafraction motion during frameless radiosurgery using Varian HyperArcTM and BrainLab ElementsTM immobilization systems. J Radiosurg SBRT. 2020;7(2):149-156.
2. 2. Ramakrishna N, Rosca F, Friesen S, Tezcanli E, Zygmanszki P, Hacker F. A clinical comparison of patient setup and intra-fraction motion using frame-based radiosurgery versus a frameless image-guided radiosurgery system for intracranial lesions. Radiother Oncol. 2010;95(1):109-15. https://doi.org/10.1016/j.radonc.2009.12.03010.1016/j.radonc.2009.12.03020116123
3. 3. Bisht RK, Natanasabapathi G, Kale SS. Quantitative analysis of Gamma Knife stereotactic radiosurgery for uveal melanoma. Polish Journal of Medical Physics and Engineering. 2020;26(2):111-117. https://doi.org/10.2478/pjmpe-2020-001210.2478/pjmpe-2020-0012
4. 4. Trifiletti DM, Chao ST, Sahgal A, Sheehan JP. Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy. A Comprehensive Guide. Springer Nature, Switzerland; 201910.1007/978-3-030-16924-4
5. 5. Mamballikalam G, Senthilkumar S, Ahamed BPM, et al. Point dose verification of Cranial Stereotactic Radiosurgery using micro Ionization Chamber and EBT3 film for 6MV FF and FFF beams in Varian TrueBeam LINAC. Polish Journal of Medical Physics and Engineering. 2020;26(3):135-142. https://doi.org/10.2478/pjmpe-2020-001510.2478/pjmpe-2020-0015