Point dose verification of Cranial Stereotactic Radiosurgery using micro Ionization Chamber and EBT3 film for 6MV FF and FFF beams in Varian TrueBeam® LINAC

Author:

Mamballikalam Gopinath1,Senthilkumar S.2,Ahamed Basith P.M.3,Inipully Rohit3,Jayadevan P.M.3,Clinto C.O.3,Jaon Bos R.C.3

Affiliation:

1. R & D, Bharathiar University , Coimbatore, Tamilnadu, India & Aster Medcity, Kochi, Kerala , India

2. Rajaji Hospital & Madurai Medical College , Madurai , Tamilnadu , India

3. Aster Medcity , Kochi , Kerala , India

Abstract

Abstract Introduction: Achieving high positional and dosimetric accuracy in small fields is very challenging due to the imbalance of charged particle equilibrium (CPE), occlusion of the primary radiation source, and overlapping penumbra regions. These factors make the choice of the detector for Stereotactic Radiosurgery (SRS) patient-specific quality assurance (PSQA) difficult. The aim of the study is to compare the suitability of EBT3 Gafchromic film against CC01 pinpoint chamber for the purpose of SRS and stereotactic Radiotherapy (SRT) dose verification. Material and Method: EBT3 Gafchromic film was calibrated against Treatment Planning System (TPS) doses (1 Gy – 35 Gy). CC01 pinpoint chamber and EBT3 film was used to verify Patient-Specific point doses of 21 intracranial lesions each planned with Static, Dynamic Conformal Arc (DCA), and Volumetric Arc Therapy (VMAT) using Varian TrueBeam Accelerator 6MV Flattening Filter (FF) and 6MV Flattening Filter Free (FFF) beams. The lesion sizes varied from 0.4 cc to 2.9 cc. The lesions were categorized into <1cc, 1cc-2cc and 2cc-3cc. Results: High variations in measured doses from TPS calculated dose were observed with small lesion volumes irrespective of the dosimeter. As the sizes decreased high uncertainty was observed in ion chamber results. CC01 was observed under-responding to film in small lesion sizes (<1cc), where nearly 50% of results under-responded in comparison with Film results. Film results were more or less consistent for static and DCA plans. Static and DCA plans were consistent passing more than 73% of the plans of the smallest lesion size category. VMAT showed very poor PSQA agreement for all three volumes (32.1% for <1cc, 14.3% for 2cc-3cc and 39.3% for 2cc-3cc). No significant difference was observed between 6MVFF and 6MVFFF beams from the chi-squared test. Conclusion: EBT3 Film was observed to be a more suitable detector for small lesion sizes less than 1cc, compared to CC01. As the volume increases, the response of CC01 and EBT3 film have no significant difference in performing PSQA for intracranial SRS/SRT. VMAT techniques for intra cranial SRS shows deviation from TPS planned dose for both EBT3 film and CC01 and should not be preferred choice of verification tools.

Publisher

Walter de Gruyter GmbH

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3