Author:
Mitrev Ludmil,van Helmond Noud,Kaddissi Georges,Awad Ahmed,Patel Kinjal,Aji Janah,Ogbara Jeffrey,Rafeq Zahi,Nagubandi Vineeth,Orr Debbie,Gaughan John,Rosenbloom Michael
Abstract
AbstractCardiac output during cardiac catheterization is often estimated using the modified Fick method (COFick). In this proof-of-concept, prospective non-randomized study carried out in a single academic healthcare centre, we examined whether replacing COFick in the Gorlin formula with Cheetah-NICOM monitor cardiac output (COCheetah) could produce an accurate and precise estimate of aortic valve area in patients with severe aortic stenosis. In twenty-six subjects, COFick and COCheetah were obtained concurrently. A spot and 3-minute running average of COCheetah was used. Bland and Altman analysis was used to derive bias, 95% limits of agreement (LOA) and confidence intervals (CI). The mean difference (bias) between AVACheetah (average) and AVAFick was 0.11 cm2 and the 95% LOA were ±0.42 cm2. The 95% CI of the bias was 0.02–0.2 cm2. The bias and 95% LOA of AVACheetah (spot value) were 0.14 ± 0.42cm2, with a 95% CI of 0.06–0.23 cm2. No proportional bias was present. AVACheetah thus appears to be a reasonably accurate measure of AVA in patients with severe aortic stenosis compared to AVAFick measured using a modified Fick CO. However, the limits of agreement were not narrow enough to consider AVACheetah and AVAFick interchangeable.
Publisher
Springer Science and Business Media LLC
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献