Abstract
Tracking unreported allergens in commercial foods can avoid acute allergic reactions. A 2-step electrochemical immunosensor was developed for the analysis of the peanut allergen Ara h 1 in a 1-h assay (<15 min hands-on time). Bare screen-printed carbon electrodes (SPCE) were used as transducers and monoclonal capture and detection antibodies were applied in a sandwich-type immunoassay. The short assay time was achieved by previously combining the target analyte and the detection antibody. Core/shell CdSe@ZnS Quantum Dots were used as electroactive label for the detection of the immunological interaction by differential pulse anodic stripping voltammetry. A linear range between 25 and 1000 ng·mL−1 (LOD = 3.5 ng·mL−1), an adequate precision of the method (Vx0 ≈ 6%), and a sensitivity of 23.0 nA·mL·ng−1·cm−2 were achieved. The immunosensor was able to detect Ara h 1 in a spiked allergen-free product down to 0.05% (m/m) of peanut. Commercial organic farming cookies and cereal and protein bars were tested to track and quantify Ara h 1. The results were validated by comparison with an ELISA kit.
Funder
Fundação para a Ciência e Tecnologia
European Union
Subject
Clinical Biochemistry,General Medicine
Cited by
16 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献