Author:
Li Junhong,Ren Wei,Fan Guoxiang,Wang Chenghao
Abstract
A square piezoelectric composite diaphragm was analyzed by the finite element method to enhance the sensitivity of a piezoelectric micromachined ultrasound transducer (pMUT). The structures of electrode and piezoelectric film were optimized and a centric electrode was designed to avoid the counteraction of stress in the centre and edges. In order to further improve the sensitivity; a pMUT with partially-etched piezoelectric film was adopted. The receive and transmit sensitivities of the pMUT were analyzed in details. The receive sensitivity of pMUT with partially-etched ZnO film is 3.3 dB or 6.8 dB higher than those with a centric and whole electrode, respectively; and the amplitude of a partially-etched ZnO film pMUT under a certain voltage is 5.5 dB and 30 dB higher than those with centric and whole electrode separately. Two pMUT-based ZnO films were fabricated by micromachining technology and their receive and transmit sensitivities were tested. The ZnO films deposited by direct current (DC) magnetron sputtering exhibit a densely packed structure with columnar crystallites. The test results show that the structure of the square diaphragm with partially-etched piezoelectric layer can significantly improve the transducer sensitivity. The receive sensitivity and transmit sensitivity are −238.35 dB (ref. 1 V/μPa) and 150.42 dB (ref. 1 μPa/V); respectively.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
29 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献