Author:
Han Zhen-Wei,Song Ke-Fei,Zhang Hong-Ji,Yu Miao,He Ling-Ping,Guo Quan-Feng,Wang Xue,Liu Yang,Chen Bo
Abstract
The radiation intensity of observed auroras in the far-ultraviolet (FUV) band varies dramatically with location for aerospace applications, requiring a photon counting imaging apparatus with a wide dynamic range. However, combining high spatial resolution imaging with high event rates is technically challenging. We developed an FUV photon counting imaging system for aurora observation. Our system mainly consists of a microchannel plate (MCP) stack readout using a wedge strip anode (WSA) with charge induction and high-speed electronics, such as a charge sensitive amplifier (CSA) and pulse shaper. Moreover, we constructed an anode readout model and a time response model for readout circuits to investigate the counting error in high counting rate applications. This system supports global rates of 500 kilo counts, 0.610 dark counts s−1 cm−2 at an ambient temperature of 300 K and 111 µm spatial resolution at 400 kilo counts s−1 (kcps). We demonstrate an obvious photon count loss at incident intensities close to the counting capacity of the system. To preserve image quality, the response time should be improved and some noise performance may be sacrificed. Finally, we also describe the correlation between counting rate and imaging resolution, which further guides the design of space observation instruments.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献