Abstract
The performance of the CdTe radiation detectors heavily relies on the method of contact preparation. A convenient research method addressing this problem is the laser-induced transient current technique. In this paper, we compare the performance of two CdTe crystals which underwent different metallization processes. We showed that appropriately designed Au/Al contacts induce much less bulk polarization than commercial Pt/In electrodes under the same working conditions and can thus provide a convenient alternative to the industry standard. The comparison was based on the monitoring of the time-dependent sensor polarization measuring transient currents excited by above-bandgap laser illumination complemented by the Am 241 gamma spectroscopy. The theoretical analysis of current waveforms and radiation spectra enabled us to determine the charge carrier mobility, mobility-lifetime products of electrons and holes, and temporal and bias dependence of the space charge formation.
Funder
Grant Agency of the Czech Republic
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
6 articles.