Swirling Flow and Wall Shear: Evaluating the BioMimics 3D Helical Centerline Stent for the Femoropopliteal Segment

Author:

Sullivan Timothy M.1ORCID,Zeller Thomas2,Nakamura Masato3,Caro Colin G.4,Lichtenberg Michael5

Affiliation:

1. Vascular/Endovascular Surgery, Minneapolis Heart Institute at Abbott Northwestern Hospital, Minneapolis, MN, USA

2. Department of Angiology, Universitäts-Herzzentrum Freiburg Bad Krozingen, Bad Krozingen, Germany

3. Division of Cardiovascular Medicine, Toho University, Ohashi Medical Center, Tokyo, Japan

4. Department of Bioengineering, Imperial College, London, UK

5. Vascular Center Klinikum Arnsberg, Arnsberg, Germany

Abstract

The BioMimics 3D self-expanding nitinol stent represents a strategy for femoropopliteal intervention that is alternative or complementary to deployment of drug-coated stents or balloons. Whereas conventional straight stents reduce arterial curvature and disturb blood flow, creating areas of low wall shear, where neointimal hyperplasia predominantly develops, the helical centerline geometry of the BioMimics 3D maintains or imparts arterial curvature, promotes laminar swirling blood flow, and elevates wall shear to protect against atherosclerosis and restenosis. In the multicenter randomized MIMICS trial, treatment of femoropopliteal disease with the BioMimics 3D (n=50) significantly improved 2-year primary patency (log-rank test p=0.05) versus a control straight stent (n=26), with no cases of clinically driven target lesion revascularization between 12 and 24 months (log-rank test p=0.03 versus controls). In geometric X-ray analysis, the BioMimics stent was significantly more effective in imparting a helical shape even when the arterial segment was moderately to severely calcified. Computational fluid dynamics analysis showed that average wall shear was significantly higher with the helical centerline stent (1.13±0.13 Pa versus 1.06±0.12 Pa, p=0.05). A 271-patient multicenter international MIMICS-2 trial and a 500-patient real-world MIMICS-3D registry are underway.

Funder

Veryan Medical

Publisher

Hindawi Limited

Subject

Cardiology and Cardiovascular Medicine

Cited by 16 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3