Deformation of the Femoropopliteal Segment

Author:

Ní Ghriallais Ríona1,Heraty Kevin2,Smouse Bob3,Burke Martin2,Gilson Paul2,Bruzzi Mark1

Affiliation:

1. Department of Biomedical Engineering, National University of Ireland, Galway, Ireland

2. Veryan Medical, Galway Technology Park, Parkmore, Galway, Ireland

3. Department of Radiology, University of Illinois College of Medicine, Peoria, IL, USA

Abstract

Purpose: To quantify the deformation behavior of the diseased femoropopliteal segment and assess the change to deformation behavior due to various stent placements. Methods: The length and curvature changes of 6 femoropopliteal segments (the right and left superficial femoral and popliteal arteries) from 3 cadavers were measured in 3-dimensional space based on rotational angiography image data in straight leg and flexed hip/knee (50°/90°) positions before and after placement of nitinol stents of varying type (EverFlex, Misago, and BioMimics 3D) and length (60, 100, and 200 mm) in different locations along the arteries. Three-dimensional centerline data were extracted for the measurements. Results: All 6 femoropopliteal cadaver segments displayed signs of peripheral artery disease. Hip/knee flexion resulted in vessel shortening and increases in the mean and maximum vessel curvatures in all cases. Location-specific results of the unstented arteries showed that magnitudes of vessel length and curvature change vary as a function of vessel length. The average shortening of the entire femoropopliteal segment due to flexion was observed at 10.7%±0.7%, which was reduced to 8.1%±0.9% after stent deployment. Average and maximum curvatures of the unstented segment increased due to flexion (average: 0.008±0.002 mm−1 to 0.019±0.006 mm−1, maximum: 0.030±0.009 mm−1 to 0.091±0.045 mm−1). After stent deployment, average and maximum curvatures of the flexed stented segments increased compared with the flexed unstented segments (average: 0.019±0.006 mm−1 to 0.022±0.004 mm−1, maximum: 0.091±0.045 mm−1 to 0.103±0.025 mm−1). The most flexurally stiff stent demonstrated the least ability to axially shorten during flexion of the leg at the knee joint. Conclusion: The deformation characteristics of the femoropopliteal segment change in the presence of a stent, with the change to the deformation behavior dependent on stent type, stent length, location, flexibility, and intrinsic centerline curvature.

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging,Surgery

Cited by 21 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3