Author:
Olms Kai,Randt Thorsten,Reimers Nils,Zander Nils,Schulz Arndt P.
Abstract
Reconstructive osteotomies for the treatment of Hallux valgus are among the most prevalent procedures in foot
and ankle surgery.
The combination of biodegradable materials with an innovative method for fixation by application of ultrasonic energy
facilitates a new bonding method for fractures or osteotomies. As clinical experience is still limited, the aim of this study
was to assess the safety and performance of the SonicPin system for fixation of Austin/Chevron osteotomies.
Chevron osteotomy was performed on 30 patients for the treatment of Hallux valgus. The used SonicPins were made from
polylactide and are selectively melted into the cancellous bone structure during insertion by ultrasonic energy. Patients
were followed for one year, which included X-ray and MRI examinations as well as evaluation of life quality by EQ-5D
(EuroQol).
The MRI after three months showed adequate bone healing in all cases and no signs of foreign body reactions, which was
again confirmed by MRI 12 months postoperatively. The bony healing after 12 months was uneventful without any signs
of foreign body reactions.
In summary, based on the low complication rate and the significant improvement in health related quality of life (EQ-5D)
reported in this study, fixation of an Austin/Chevron osteotomy with a SonicPin for treatment of Hallux valgus can be
considered to be safe and efficient over the short term.
Level of Clinical Evidence: Therapeutic Level III.
Publisher
Bentham Science Publishers Ltd.
Reference43 articles.
1. Austin DW, Leventen EO.
A new osteotomy for hallux valgus: a horizontally directed "V" displacement ostoetomy of the metatarsal head for hallux valgus and primus cavus.
Clin Orthop Relat Res
1981;
157
: 25-30.
2. Sorensen MD, Hyer CF.
Metatarsus primus varus correction: the osteotomies.
Clin Podiatr Med Surg
2009;
26
(3)
: 409-25.
3. Crosby LA, Bozarth GR.
Fixation comparison for chevron osteotomies.
Foot Ankle Int
1998;
19
(9462912)
: 41-3.
4. Hanft JR, Kashuk KB, Bonner AC, Toney M, Schabler J.
Rigid internal fixation of the Austin/Chevron osteotomy with Herbert screw fixation: a retrospective study.
J Foot Surg
1992;
31
(5)
:
512-8.
5. Barca F, Busa R.
Austin/chevron osteotomy fixed with bioabsorbable poly-L-lactic acid single screw.
J Foot Ankle Surg
1997;
36
(1)
: 15-20.
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献