Affiliation:
1. Shanghai Institute of Optics and Fine Mechanics
2. China-Russian Belt and Road Joint Laboratory on Laser Science
Abstract
The lateral distortion of a surface measuring Fizeau interferometer may cause distorted image features in the lateral direction, as well as the surface form error in the axial direction (which is a source of the retrace error). Traditional method for lateral distortion measurement requires a high-accuracy calibration plate featuring a grid pattern. Such a calibration plate is not always available, especially when the required accuracy of the grid pattern comes to the order of sub-micrometer or even nanometer level. To remove the dependence on the plate accuracy, we propose a self-calibration method for the measurement and correction of lateral distortion in Fizeau interferometer. The self-calibration technique may separate the lateral distortion and the geometric error of the calibration plate. This method is verified using a 108-mm-aperture Fizeau interferometer. The experiments show that the form measurement error of a surface tilted at approximately 5° and 16° can be reduced from 92 nm to 43 nm and from 251 nm to 144 nm (peak-to-valley value), respectively, after the distortion correction.
Funder
National Key Research and Development Program of China
International Partnership Program of Chinese Academy of Sciences
Subject
Atomic and Molecular Physics, and Optics
Cited by
4 articles.
订阅此论文施引文献