Feasibility of Automated Gonioscopy Imaging in Clinical Practice

Author:

Barbour-Hastie Catriona12,Deol Sundeep S.12,Peroni Andrea3,Gillan Stewart4,Trucco Emanuele3,Tatham Andrew J.12

Affiliation:

1. Princess Alexandra Eye Pavilion, NHS Lothian and Centre for Clinical Brain Sciences

2. University of Edinburgh, Edinburgh

3. School of Science and Engineering, University of Dundee

4. Department of Ophthalmology, Ninewells Hospital, NHS Tayside, Dundee, UK

Abstract

Précis: Automated gonioscopy provided good-quality images of the anterior chamber angle. There was a short learning curve for operators, and the examination was well tolerated by patients. Patients expressed a preference for automated gonioscopy compared with traditional gonioscopy. Purpose: The purpose of this study was to assess the feasibility of using a desktop automated gonioscopy camera in glaucoma clinics by examining patient tolerability, ease of use, and image quality and comparing patient preference compared with traditional gonioscopy. Patients and Methods: A prospective study was conducted in a university hospital clinic. Traditional gonioscopy was performed followed by imaging of the iridocorneal angle (ICA) using the Nidek GS-1 camera by 2 glaucoma specialists. Participants were asked to rate the comfort of automated gonioscopy and which method they preferred. The clinicians graded the ease of acquisition for each patient, and the image quality was reviewed by a grader. Results: Forty-three eyes of 25 participants were included. Sixty-eight percent of participants viewed automated gonioscopy as “extremely comfortable,” and the remainder reported it “comfortable”. Forty percent preferred automated gonioscopy compared with traditional gonioscopy, while 52% were equivocal. Clinicians scored 32% of participants as “somewhat difficult” to the image. In 46% of eyes, good-quality photographs were obtained for 360 degrees of the ICA. Only 1 eye had no parts of the ICA clearly visible. Seventy-four percent of eyes had at least half of the ICA clearly visible in all 4 quadrants. Conclusion: Automated gonioscopy provided good-quality images of the ICA for most patients. It was often not possible to image the entire 360 degrees at the first attempt, but the examination was comfortable for patients, and only 8% preferred traditional gonioscopy to the automated photographic examination.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Ophthalmology

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3