Effect of bending direction on the mechanical behaviour of 3.5 mm String-of-Pearls and Limited Contact Dynamic Compression Plate constructs

Author:

Benamou J.,Demianiuk R. M.,Rutherford S.,Beckett C.,Ness M. G.,Haut R. C.,Déjardin L. M.

Abstract

SummaryObjective: To compare the bending properties of String-of-Pearls® (SOP) and Limited Contact Dynamic Compression Plate® (LCDCP) constructs in orthogonal bending directions.Methods: 3.5 mm SOP and LC-DCP plates were fixed to a bone model simulating a comminuted tibial fracture. Specimens were non-destructively tested in both mediolateral and craniocaudal bending for 10 cycles. Bending stiffness and total angular deformation were compared using parametric analyses (p <0.05).Results: For both constructs, stiffness was significantly less when bending moments were applied against the thickness of the plates (mediolateral bending) than against the width (craniocaudal bending). When compared to the mediolateral plane, bending constructs in the craniocaudal plane resulted in a 49% (SOP group) and 370% (LC-DCP group) increase in stiffness (p <0.001). Mediolateral bending stiffness was significantly greater in the SOP than the LC-DCP constructs. Conversely, in craniocaudal bending, SOP constructs stiffness was significantly less than that of the LC-DCP constructs. The differences between the two constructs in total angular deformation had an identical pattern of significance.Clinical significance: This study found that SOP showed less variability between the orthogonal bending directions than LC-DCP in a comminuted fracture model, and also described the bi-planar bending behaviour of both constructs. Although not exhibiting identical bending properties in both planes, SOP constructs had a more homogenous bending behaviour in orthogonal loading directions. The difference between the SOP with a circular cross sectional shape compared to the rectangular shape of standard plates is probably responsible for this difference.

Publisher

Georg Thieme Verlag KG

Subject

General Veterinary,Animal Science and Zoology

Reference34 articles.

1. Johnson AL, Houlton JEF, Vannini R. AO principles of fracture management in the dog and cat. Davos Platz, Switzerland: AO Publishing; 2005

2. Area Moment of Inertia for Comparison of Implant Cross-Sectional Geometry and Bending Stiffness

3. Standard specification and test method for metallic bone plates, ASTM F382-99 (re-approved 2008) e1. Annual Book of ASTM standards. West Conshohocken, PA: ASTM; 2008

4. Mechanical properties of 18 different AO bone plates and the clamp-rod internal fixation system tested on a gap model construct

5. Single cycle to failure in bending of three standard and five locking plates and plate constructs

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献