Tracking the same fast-LGSO crystals by changing surface treatments for better coincidence timing resolution in PET

Author:

Kiyokawa Miho,Kang Han GyuORCID,Yamaya Taiga

Abstract

Abstract Achieving fast coincidence timing resolution (CTR) is an important issue in clinical time-of-flight positron emission tomography (TOF-PET) to improve the reconstructed image quality. One of the major factors affecting the CTR is the crystal surface treatment, which is often parameterized as surface roughness. However, previous studies on the crystal surface treatment optimization had two limitations of crystal-by-crystal variation and worse CTR over 200 ps. Here, we report the effects of the crystal surface treatment on the performance of a 20 mm long fast-LGSO crystal based TOF detector by tracking the same crystals in the sub-180 ps CTR regime. The light collection efficiency (LCE), energy resolution (ER) and CTR of the TOF detector were evaluated with six different crystal surface treatments of chemically polished (C.P), C.P half side roughened (1/2S) treatment, and then the C.P one side roughened (1S) treatment, mechanically polished (M.P) treatment, M.P 1/2S treatment, and M.P 1S treatment. The four lateral surfaces of each crystal were wrapped by using enhanced specular reflector film while the top surface was covered by using Teflon tape. The bottom surface of the crystal was optically coupled to a silicon photomultiplier. The timing and energy signals were extracted by using a custom-made high-frequency readout circuit, and then digitized by using a waveform digitizer. All the experimental conditions were same except the crystal surface treatment. Among the six different crystal surface treatments, the M.P 1S would be the optimal crystal surface treatment which balanced enhancements in the CTR (165 ± 3 ps) and ER (10.5 ± 0.5%). Unlike the M.P 1S, the C.P 1S did not enhance the CTR and ER. Hence, the C.P without roughening would be the second-best optimal crystal surface treatment which balanced the CTR (169 ± 3 ps) and ER (10.5 ± 0.5%).

Funder

Japan Society for the Promotion of Science

Nakatani Foundation for Advancement of Measuring Technologies in Biomedical Engineering

National Institutes for Quantum and Radiological Science and Technology

Publisher

IOP Publishing

Subject

General Nursing

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3