We present the performance of a digital phoswich positron emission tomography (PET) detector, composed by layers of pixilated scintillator arrays, read out by solid state light detectors and an application specific integrated circuit (ASIC). We investigated the use of integrated charge from the scintillation pulses along with time-over-threshold (ToT) to determine the layer of interaction (DOI) in the scintillator. Simulations were performed to assess the effectiveness of the ToT measurements for separating the scintillator events and identifying cross-layer-crystal-scatter (CLCS) events. These simulations indicate that ToT and charge integration from such a detector provide sufficient information to determine the layer of interaction. To demonstrate this in practice, we used a pair of prototype LYSO/BGO detectors. One detector consisted of a 19 × 19 array of 7 mm long LYSO crystals (1.36 mm pitch) coupled to a 16 × 16 array of 8 mm long BGO crystals (1.63 mm pitch). The other detector was similar except the LYSO crystal pitch was 1.63 mm. These detectors were coupled to an 8 × 8 multi-pixel photon counter mounted on a PETsys TOFPET2 ASIC. This high performance ASIC provided digital readout of the integrated charge and ToT from these detectors. We present a method to separate the events from the two scintillator layers using the ToT, and also investigate the performance of this detector. All the crystals within the proposed detector were clearly resolved, and the peak to valley ratio was 11.8 ± 4.0 and 10.1 ± 2.9 for the LYSO and BGO flood images. The measured energy resolution was 9.9% ± 1.3% and 28.5% ± 5.0% respectively for the LYSO and BGO crystals in the phoswich layers. The timing resolution between the LYSO–LYSO, LYSO–BGO and BGO–BGO coincidences was 468 ps, 1.33 ns and 2.14 ns respectively. Results show ToT can be used to identify the crystal layer where events occurred and also identify and reject the majority of CLCS events between layers.
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Cited by
16 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Design and characterization of a hybrid PET detector with DOI capability;Medical Physics;2024-07-20
2. A continuous depth encoding PET detector using side readout of dual-layer GAGG crystals with SiPM array;Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment;2024-02
3. Timing estimation of the exponentiated energy-weighted average for crosshair light sharing TOF-DOI PET detector;Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment;2024-02
4. A multi-resolution TOF-DOI detector for human brain dedicated PET scanner;Physics in Medicine & Biology;2024-01-21
5. An LYSO and BGO checkerboard pattern detector with inter-crystal scatter identification capability;2023 IEEE Nuclear Science Symposium, Medical Imaging Conference and International Symposium on Room-Temperature Semiconductor Detectors (NSS MIC RTSD);2023-11-04