Design and characterization of a hybrid PET detector with DOI capability

Author:

He Wen12,Zhao Yangyang1,Zeng Honghao13,Huang Wenjie1,Yang Hang1,Zhao Xin1,Wang Qiang4,Wang Lu4,Niu Ming1,Zhang Lei25,Ren Qiushi12,Gu Zheng12

Affiliation:

1. Institute of Biomedical Engineering Shenzhen Bay Laboratory Shenzhen China

2. Peking University Shenzhen Graduate School Shenzhen China

3. Ciechanover Institute of Precision and Regenerative Medicine School of Medicine The Chinese University of Hong Kong Shenzhen China

4. 26th Institute of China Electronics Technology Group Corporation Chongqing China

5. Institute of Cancer Research Shenzhen Bay Laboratory Shenzhen China

Abstract

AbstractBackgroundMonolithic or semi‐monolithic detectors are attractive for positron emission tomography (PET) scanners with depth‐of‐interaction (DOI) capability. However, they often require complicated calibrations to determine the interaction positions of gamma photons.PurposeWe introduce a novel hybrid detector design that combines pixelated and semi‐monolithic elements to achieve DOI capability while simplifying the calibrations for positioning.MethodsA prototype detector with eight hybrid lutetium–yttrium oxyorthosilicate (LYSO) layers having dimensions of 25.8 × 12.9 × 15 mm3 was constructed. The energy‐weighted and energy‐squared weighted averages were used for estimating the x‐ (pixelated direction) and y‐positions (non‐pixelated direction). Pseudo‐pixels were defined as discrete areas on the flood image based on the crystal look‐up table (LUT). The intrinsic spatial resolutions in the pixelated and non‐pixelated directions were measured. The ratio of the maximum to the sum of the multipixel photon counter (MPPC) signals was used to estimate the DOI positions. The coincidence timing resolution (CTR) was measured using the average and energy‐weighted average of the earliest n time stamps. Two energy windows of 250–700 and 400–600 keV were applied for the measurements.ResultsThe pattern of the flood images showed discrete event clusters, demonstrating that simple calibrations for determining the x‐ and y‐positions of events could be achieved. Under 400–600 keV energy window, the average intrinsic spatial resolutions were 1.15 and 1.34 mm for the pixelated and non‐pixelated directions; the average DOI resolution of the second row of pseudo‐pixels was 5.1 mm in full width at half maximum (FWHM); when using the energy‐weighted average of the earliest four‐time stamps, the best CTR of 350 ps was achieved. Applying a broader energy window of 250–700 keV only slightly degrades the DOI resolution while maintaining the intrinsic resolution; the best CTR degrades to 410 ps.ConclusionsThe proposed hybrid detector concept was verified, and a prototype detector showed high performance for 3D positioning and timing resolution. The novel detector concept shows promise for preclinical and clinical PET scanners with DOI capability.

Funder

National Natural Science Foundation of China

Publisher

Wiley

Reference40 articles.