Author:
Whiteley Greg S.,Derry Chris,Glasbey Trevor,Fahey Paul
Abstract
OBJECTIVETo investigate the reliability of commercial ATP bioluminometers and to document precision and variability measurements using known and quantitated standard materials.METHODSFour commercially branded ATP bioluminometers and their consumables were subjected to a series of controlled studies with quantitated materials in multiple repetitions of dilution series. The individual dilutions were applied directly to ATP swabs. To assess precision and reproducibility, each dilution step was tested in triplicate or quadruplicate and the RLU reading from each test point was recorded. Results across the multiple dilution series were normalized using the coefficient of variation.RESULTSThe results for pure ATP and bacterial ATP from suspensions ofStaphylococcus epidermidisandPseudomonas aeruginosaare presented graphically. The data indicate that precision and reproducibility are poor across all brands tested. Standard deviation was as high as 50% of the mean for all brands, and in the field users are not provided any indication of this level of imprecision.CONCLUSIONSThe variability of commercial ATP bioluminometers and their consumables is unacceptably high with the current technical configuration. The advantage of speed of response is undermined by instrument imprecision expressed in the numerical scale of relative light units (RLU).Infect Control Hosp Epidemiol2015;00(0):1–6
Publisher
Cambridge University Press (CUP)
Subject
Infectious Diseases,Microbiology (medical),Epidemiology