Bolstad Nils,Warren David J.,Bjerner Johan,Kravdal Gunnhild,Schwettmann Lutz,Olsen Kari H.,Rustad Pål,Nustad Kjell
Abstract
AbstractHeterophilic antibodies are still an important source of interference in immunoassays. We have conducted a screening study for interference in a panel of commercially available assays using two sera known to contain high titer Fc-reactive heterophilic antibodies.The sera were distributed to laboratories participating in the Nordic External Quality Assessment cooperation (EQANord). Duplicate samples pre-blocked with aggregated murine monoclonal MAK33 were also supplied. Discrepancies (>50%) between the results for native and blocked samples were used to classify the tested assays as susceptible to interference. A total of 170 different assay kits covering 91 analytes were tested.We found that 21 assays, covering 19 different analytes, were susceptible to interference from the heterophilic antibodies in the two sera. Many of these are clinically and commercially important assays. Some of the false results were grossly elevated and could have been detrimental to patient care in a clinical setting.Heterophilic antibodies with Fc-reactivity remain a threat. A more widespread use of antibody fragments and aggregated immunoglobulin could potentially improve the heterophilic antibody resistance of assays intended for clinical use.