Mori A.,Oda H.,Onozawa E.,Shono S.,Takahashi T.,Yamashita S.,Fujimoto H.,Sako T.
Abstract
Abstract
This study evaluated the accuracy and reproducibility of a human portable blood glucose meter (PBGM) for canine and feline whole blood. Reference plasma glucose values (RPGV) were concurrently measured using glucose oxidation methods. Fifteen healthy dogs and 6 healthy cats were used for blood sampling. Blood glucose concentrations and hematocrits were adjusted using pooled blood samples for our targeted values. A positive correlation between the PBGM and RPGV was found for both dogs (y = 0.877, x = −24.38, r = 0.9982, n = 73) and cats (y = 1.048, x = −27.06, r = 0.9984, n = 69). Acceptable results were obtained in error grid analysis between PBGM and RPGV in both dogs and cats; 100% of these results were within zones A and B. Following ISO recommendations, a PBGM is considered accurate if 95% of the measurements are within ± 15 mg/dl of the RPGV when the glucose concentration is <100 mg/dl and within ±15% when it is ≥100 mg/dl; however, small numbers of samples were observed inside the acceptable limits for both dogs (11%, 8 of 73 dogs) and cats (39%, 27 of 69 cats). Blood samples with high hematocrits induced lower whole blood glucose values measured by the PBGM than RPGV under hypoglycemic, normoglycemic, and hyperglycemic conditions in both dogs and cats. Therefore, this device is not clinically useful in dogs and cats. New PBGMs which automatically compensate for the hematocrit should be developed in veterinary practice.
Subject
General Veterinary,General Medicine
Reference30 articles.
1. Evaluation of artificial pancreas technology for continuous blood glucose monitoring in dogs;Mori;J Artif Organs,2011
2. System accuracy evaluation of blood glucose monitoring systems for self - monitoring of blood glucose according to DIN EN ISO;Freckmann;Diabetes Sci Technol,2012
3. DJ BP DM Understanding error grid analysis;Cox;Diabetes Care,1997
4. Carbohydrates In Textbook of Molecular Diagnostics thed CA ER eds St MO;Sacks;Clinical Chemistry,2006
5. System accuracy evaluation of blood glucose monitoring systems for self - monitoring of blood glucose according to DIN EN ISO;Freckmann;Diabetes Sci Technol,2012
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献